Analytical Instruments
DOG-208FA Dissolved Oxygen Electrode
Steam-sterilisable, 130 °C

DOG-208FA technical data
Measurement
- Operating range
- 0.01–20 ppm
- Response time
- About 10 s to 95 % (rising)
- Response current in air
- About 60 nA
- Response current in nitrogen
- < 1 % of the in-air value
- Stability
- Drift < 3 % per week at constant temperature and oxygen partial pressure
- Stirring effect
- < 3 % in water at room temperature
- Temperature coefficient
- About 3 % at room temperature
Construction
- Body material
- Stainless steel
- Membrane
- Fluoroplastic / silicone rubber / stainless wire mesh composite
- Cathode
- 0.5 mm platinum wire
- Anode
- 0.3 mm silver wire
- Electrolyte
- Chloride solution
- Temperature sensor
- 22 kΩ NTC thermistor at room temperature
- Insertion diameter
- 12 mm
- Insertion length
- 120 mm, 280 mm or 450 mm
Values are for the standard configuration. Ranges, threads, probe lengths and output types outside this table are available on request — the datasheet revision for a custom build is issued with the quotation.
Request the PDF datasheet →Analytical Instruments
DOG-208FADissolved Oxygen Electrode
Steam-sterilisable, 130 °C
Polarographic dissolved oxygen electrode that withstands 130 °C steam sterilisation, with a self-pressure-balancing design and an imported BJC membrane. 0.01–20 ppm range, 10 s response, 12 mm insertion diameter in 120, 280 or 450 mm lengths.
Engineering reply · Evaluation samples available · Custom ranges built to order
- Withstands 130 °C steam sterilisation
- Self-pressure-balancing — membrane stays seated
- Imported BJC membrane head
- Built-in 22 kΩ NTC temperature sensor
- Insertion lengths 120, 280 or 450 mm
- Response about 10 s to 95 %
How it works
The DOG-208FA is built for bioreactors and fermenters, where the electrode has to survive repeated steam-in-place cycles and still hold calibration. The pressure-balancing construction prevents the membrane from being pushed off its seat when the vessel is pressurised, which is the usual failure mode for an ordinary electrode in a fermenter. The membrane head is an imported BJC type, and the body is stainless with a built-in NTC thermistor for temperature compensation, so the transmitter does not need a separate probe in the same vessel.
Ordering notes
- Pair with the DOG200 transmitter, or any polarographic DO input with 22 kΩ NTC compensation.
- Membrane and electrolyte are consumables; keep spares on site for sterilisation cycles.
Typical applications
- Fermenters and bioreactors with SIP
- Cell culture vessels
- Pharmaceutical and food fermentation
Not certain DOG-208FA is the right unit?
Send the medium, range, process connection and the signal your control system accepts. We will confirm the model — or tell you which one fits better — before you commit to an order.
Same family
Other analytical instruments

Dissolved Oxygen Meter
Panel-mount, 96 × 96
Panel-mount dissolved oxygen transmitter reading 0–19.99 mg/L or 0–199.9 %, with automatic temperature compensation, isolated 4–20 mA output and dual relay control. 96 × 96 mm DIN case, 230 V AC supply.
View specifications
pH / ORP Controller
Two analogue outputs, RS485
pH and ORP controller accepting a range of analogue electrodes through a built-in A/D stage. −2 to 16 pH, −2000 to 2000 mV, ±1 % FS, two 4–20 mA outputs, five relays, Modbus RTU, ABS housing, IP65.
View specifications