DOG-208FA Dissolved Oxygen Electrode

Steam-sterilisable, 130 °C

Request a quote

DOG-208FA technical data

Measurement

Operating range
0.01–20 ppm
Response time
About 10 s to 95 % (rising)
Response current in air
About 60 nA
Response current in nitrogen
< 1 % of the in-air value
Stability
Drift < 3 % per week at constant temperature and oxygen partial pressure
Stirring effect
< 3 % in water at room temperature
Temperature coefficient
About 3 % at room temperature

Construction

Body material
Stainless steel
Membrane
Fluoroplastic / silicone rubber / stainless wire mesh composite
Cathode
0.5 mm platinum wire
Anode
0.3 mm silver wire
Electrolyte
Chloride solution
Temperature sensor
22 kΩ NTC thermistor at room temperature
Insertion diameter
12 mm
Insertion length
120 mm, 280 mm or 450 mm

Values are for the standard configuration. Ranges, threads, probe lengths and output types outside this table are available on request — the datasheet revision for a custom build is issued with the quotation.

Request the PDF datasheet →

Analytical Instruments

DOG-208FADissolved Oxygen Electrode

Steam-sterilisable, 130 °C

Polarographic dissolved oxygen electrode that withstands 130 °C steam sterilisation, with a self-pressure-balancing design and an imported BJC membrane. 0.01–20 ppm range, 10 s response, 12 mm insertion diameter in 120, 280 or 450 mm lengths.

Engineering reply · Evaluation samples available · Custom ranges built to order

  • Withstands 130 °C steam sterilisation
  • Self-pressure-balancing — membrane stays seated
  • Imported BJC membrane head
  • Built-in 22 kΩ NTC temperature sensor
  • Insertion lengths 120, 280 or 450 mm
  • Response about 10 s to 95 %

How it works

The DOG-208FA is built for bioreactors and fermenters, where the electrode has to survive repeated steam-in-place cycles and still hold calibration. The pressure-balancing construction prevents the membrane from being pushed off its seat when the vessel is pressurised, which is the usual failure mode for an ordinary electrode in a fermenter. The membrane head is an imported BJC type, and the body is stainless with a built-in NTC thermistor for temperature compensation, so the transmitter does not need a separate probe in the same vessel.

Ordering notes

  • Pair with the DOG200 transmitter, or any polarographic DO input with 22 kΩ NTC compensation.
  • Membrane and electrolyte are consumables; keep spares on site for sterilisation cycles.

Typical applications

  • Fermenters and bioreactors with SIP
  • Cell culture vessels
  • Pharmaceutical and food fermentation

Not certain DOG-208FA is the right unit?

Send the medium, range, process connection and the signal your control system accepts. We will confirm the model — or tell you which one fits better — before you commit to an order.