Analytical Instruments
PHG-2091Pro pH / ORP Controller
Two analogue outputs, RS485

PHG-2091Pro technical data
Measurement
- pH range
- −2 to 16 pH
- ORP range
- −2000 to 2000 mV
- Temperature range
- −30 to +130 °C
- Accuracy
- ±1 % FS; ±0.5 °C
- Temperature compensation
- Pt1000 / NTC10K
Electrical
- Supply
- AC 90–260 V, 50/60 Hz; or DC 24 V
- Power consumption
- 4 W
- Analogue outputs
- 2 × 4–20 mA
- Relays
- 5 A at 250 V AC; 5 A at 30 V DC
- Communications
- Modbus RTU over RS485
Mechanical
- Housing
- ABS plastic
- Dimensions
- 98.2 × 98.2 × 128.3 mm
- Weight
- 0.9 kg
- Ingress protection
- IP65
Compatible electrodes
- pH
- pH8012, pH8012F, pH8010, pH8010F
- ORP
- ORP8083
Values are for the standard configuration. Ranges, threads, probe lengths and output types outside this table are available on request — the datasheet revision for a custom build is issued with the quotation.
Request the PDF datasheet →Analytical Instruments
PHG-2091PropH / ORP Controller
Two analogue outputs, RS485
pH and ORP controller accepting a range of analogue electrodes through a built-in A/D stage. −2 to 16 pH, −2000 to 2000 mV, ±1 % FS, two 4–20 mA outputs, five relays, Modbus RTU, ABS housing, IP65.
Engineering reply · Evaluation samples available · Custom ranges built to order
- Accepts most analogue pH / ORP electrodes
- Two isolated 4–20 mA outputs
- Five relay outputs plus Modbus RTU
- Solid-electrolyte reference resists blocking
- Signal cable runs over 40 m without interference
- IP65 ABS housing
How it works
The PHG-2091Pro is a pH/ORP transmitter and controller that does not lock you into one electrode. A built-in A/D stage accepts the high-impedance analogue output of most pH and ORP probes, so the instrument can be brought into service with the electrode already on the vessel. Two isolated 4–20 mA outputs carry pH and temperature separately, five relays handle dosing and alarms, and Modbus RTU over RS485 reports everything to a supervisory system. The paired electrodes use a solid-electrolyte reference with a large-area PTFE junction, which resists the blocking that stops conventional references in dirty or high-solids service.
Typical applications
- Wastewater neutralisation and dosing control
- Power plant cycle chemistry
- Chemical and metallurgical process control
- Pharmaceutical and food process water
- Municipal drinking water treatment
Not certain PHG-2091Pro is the right unit?
Send the medium, range, process connection and the signal your control system accepts. We will confirm the model — or tell you which one fits better — before you commit to an order.
Same family
Other analytical instruments

Dissolved Oxygen Meter
Panel-mount, 96 × 96
Panel-mount dissolved oxygen transmitter reading 0–19.99 mg/L or 0–199.9 %, with automatic temperature compensation, isolated 4–20 mA output and dual relay control. 96 × 96 mm DIN case, 230 V AC supply.
View specifications
Dissolved Oxygen Electrode
Steam-sterilisable, 130 °C
Polarographic dissolved oxygen electrode that withstands 130 °C steam sterilisation, with a self-pressure-balancing design and an imported BJC membrane. 0.01–20 ppm range, 10 s response, 12 mm insertion diameter in 120, 280 or 450 mm lengths.
View specifications