HSM-DOG200 Dissolved Oxygen Meter

Panel-mount, 96 × 96

Request a quote

HSM-DOG200 technical data

Measurement

Measuring range
0.00–19.99 mg/L; 0.0–199.9 %
Resolution
0.01 mg/L; 0.01 %
Accuracy
±1.5 % FS
Control range
0.00–19.99 mg/L; 0.0–199.9 %
Temperature compensation
0–65 °C

Electrical

Supply voltage
230 V AC ±10 %, 50/60 Hz
Output
Isolated 4–20 mA
Output load
Max. 500 Ω
Control output
ON/OFF relay contacts
Relay rating
5 A at 230 V AC; 10 A at 115 V AC
Isolation
500 V DC minimum to earth

Mechanical and environment

Dimensions
96 × 96 × 115 mm
Panel cut-out
91 × 91 mm
Weight
1 kg
Ambient temperature
5–55 °C
Relative humidity
≤ 80 %
Magnetic environment
No strong field beyond the earth's

Values are for the standard configuration. Ranges, threads, probe lengths and output types outside this table are available on request — the datasheet revision for a custom build is issued with the quotation.

Request the PDF datasheet →

Analytical Instruments

HSM-DOG200Dissolved Oxygen Meter

Panel-mount, 96 × 96

Panel-mount dissolved oxygen transmitter reading 0–19.99 mg/L or 0–199.9 %, with automatic temperature compensation, isolated 4–20 mA output and dual relay control. 96 × 96 mm DIN case, 230 V AC supply.

Engineering reply · Evaluation samples available · Custom ranges built to order

  • Reads 0–19.99 mg/L or 0–199.9 % saturation
  • Automatic temperature compensation 0–65 °C
  • Isolated 4–20 mA output
  • Two ON/OFF relay outputs
  • Standard 96 × 96 mm panel cut-out
  • Ten-year parameter retention without backup battery

How it works

The DOG200 drives a polarographic dissolved oxygen electrode and displays the result in mg/L or percent saturation. It handles the compensation aeration processes actually need: temperature compensation from 0 to 65 °C, and an isolated 4–20 mA output so the signal reaching the PLC does not carry the plant's earth noise. Two relays with adjustable set-points can run a blower or an alarm directly. The 96 × 96 mm cut-out is the standard panel instrument size, which means it drops into an existing cut-out.

Typical applications

  • Activated sludge aeration control
  • Aquaculture and fish-farm oxygen monitoring
  • Boiler feedwater deaeration
  • Beverage and brewing process water